|               Solvent Recycler | 
          
            |  | 
          
            | Description | 
          
          
            | Intelligent mobile phase recycling is accomplished by  placing an electronically actuated Recycle Valve in the fluid path between the  HPLC detector and the solvent waste reservoir. The signal processor in the  recycler controls the valve and monitors the detector signal to determine when  peaks appear and when contaminants are passing through the detector flow cell.  When a container is detected, the Recycle Valve is diverted to the solvent  container. When there is no contaminant detected, the Recycle Valve will bring  clean solvent back to the mobile phase reservoir for reuse. | 
          
            | Using this  solvent recycler, you can immediately regain your initial investment and enjoy  substantial saving of expense and time for years. In addition, this instrument  can reduce pollution toward environment, making more efforts to environmental  protection. | 
          
            | Features | 
          
            | - Recycle more than 90% of mobile  phase. | 
          
            | - Manual and automatic program  options allow user to control recycling and discarding of solvent. | 
          
            | - The purity of mobile phase can  be controlled precisely by the 4-degree threshold switch. | 
          
            | - Time delay of the valve switch  can be set from 10 to 90 seconds. | 
          
          
            | - The selectable three-range  input degrees match various HPLC detectors. | 
          
            | - The polarity and the changing  level of detecting signals are showed on the LED display. | 
          
            | - Auto zero setting. | 
          
          
            |  | 
          
            | Specifications | 
          
            | Range: Range of this instrument  (10mV, 100mV, or1V) matches the HPLC detectors. | 
          
            | Threshold: | 
          
            |  When the detector’s output signal exceeds the selected Threshold  value (0.1%, 0.5%, 1%, 5% of full scale), it represents it is the peak of  contaminants. The Recycle Valve then sends the contaminated effluent to waste  container. The Recycle Valve returns to the recycle position after the detector  signal falls below the Threshold value and the Time Delay has elapsed. | 
          
            | Time Delay: | 
          
            |  It holds the Recycle Valve in the waste position for the specified  amount of time (10, 20…, until 90 seconds) after the detector signal drops  below the Threshold valve. The Time Delay accounts for tailing peaks and  plumbing volume between the detector and the Recycle Valve. With precise  control of these three parameters, the solvent recycling can be accurately and  reliably operated without interference. | 
          
            |  | 
          
            |  | 
          
            | 
              
                | Model | Solvent Recycler |  
                | Recycle Valve | 2 Position Electronically Actuated All wetted parts    are Peek and TFE Fittings    are 1/16”, 1/4”-28 Flange-Free
 |  
                | Max Flow | 50mL/min (As supplied with 1.5M    1/16” Tubing on each port )
 |  
                | Analog Input | 10mV, 100mV and 1V |  
                | Display | Red and    Green LED |  
                | Zero Setting | Automatic Zero    Setting |  
                | Remote Input | Switch or TTL    Signal |  
                | Power Requirement | 13.5Volts DC/500mA |  
                | Dimensions | 7.28” × 6” × 2.7” (L × W ×    H) (18.5cm × 14.2cm × 6.28cm) |  
                | Weight | 2.21bs    (1.0Kg) |  | 
          
            |  | 
          
            |  |